PosiTector® 6000
PA2
Instruction
Manual Addendum
This addendum to instruction manual
v7.0 is intended to provide support for the SSPC-PA2
feature found in all PosiTector 6000 Advanced
models with serial numbers greater than 700000. PosiTector 6000 gages with serial numbers under 700000 - click here
Step
1 - Begin a PA2 analysis.
| Select the Memory option from
the gage menu as shown. |
Next, select the PA2 option
from the Memory menu |
|
|
|
Review the 6 parameters.
To adjust
a parameter, use the up and down
buttons to move the highlighting bar to the appropriate
parameter, then use the (+) and
(-) buttons to adjust that parameter. |
 |
The
6 adjustable parameters are...
MIN:
the specified minimum thickness value
for the coating system. If only one target thickness
is specified, set the MIN to that value.
Adjust
the displayed value down (-) or
up (+). Alternatively, measure
a coating with a thickness close to the required
value and make final adjustments with the buttons.
|
 |
MAX:
the specified maximum thickness value
for the structure. This value should be that
thickness above which adverse performance characteristics
appear, such as mudcracking, embrittlement, solvent
retention, loss of adhesion, cohesive splitting,
etc.
If a
maximum thickness value is not specified, set
the MAX value to 0.
Adjust
the displayed value down (-) or
up (+). Alternatively, measure
a coating with a thickness close to the required
value and make final adjustments with the buttons.
|
 |
#/spot:
the minimum number of individual gage readings
required per spot. PA2 suggests a minimum
of 3 readings be taken.
Adjust
the displayed value down (-) or
up (+). |
 |
spots/batch:
the minimum number of spots (sub-batches) required
per batch. PA2 suggests a minimum of 5 spots
be measured.
Adjust the displayed
value down (-) or up (+). |
|
In any measurement
there is a certain level of uncertainty. Two people
using the same gage will not necessarily record the
exact same number for a given spot measurement using
the same 4cm (1.5inch) diameter circle. To allow for
this natural fluctuation, an individual spot measurement
is permitted to be below the specified minimum thickness
as long as other spots in the batch area are high enough
to make the average thickness meet or exceed the specified
minimum thickness. Similar reasoning applies to maximum
thickness.
%
MIN:
PA2 specifies that no single spot measurement
be less than 80% of the specified MIN thickness.
The PosiTector 6000 allows this value to be changed
if desired.
Adjust the displayed
value down (-) or up (+). |
 |
%
MAX:
PA2 specifies that no single spot measurement
be more than 120% of the specified MAX thickness.
The PosiTector 6000 allows this value to be changed
if desired.
If the MAX value
above has been set to 0, this % calculation is
made to the MIN value.
Adjust the displayed
value down (-) or up (+). |
 |
To discard these changes and return to the main menu, select Cancel.
To accept these values
and begin a new PA2 analysis, select NEW.
If a PA2 analysis was already begun before entering
this set-up, selecting NEW will close/save
the existing PA2 batch and begin a new one with
a batch name containing the next higher number.
|
|
If NEW is
selected, a PA2 icon will appear on the gage's
display as shown.
Step
3 - Take measurements.
Begin taking
measurements in the first SPOT location. With each
measurement the following occurs...
-
The gage
BEEPS twice, and the resultant measurement value
is displayed in large numerals at the bottom of
the display.
-
The SPOT
information is updated
- the total number of individual
measurements taken in the current spot so far.
- the average of all measurements
in the spot. This is the PA2 spot measurement
value.
- a Pass ( )
or Fail ( )
determination is made for the SPOT based on the
user-specified
parameters
-
The BATCH
information is updated
- the total number of SPOTS in
the BATCH
- the average of the spot values
i.e. the "average of the averages".
- a Pass ( )
or Fail ( )
determination is made for the BATCH.
-
If all
PA2 criteria from Step
2 have been met, the pass symbol ( )
will appear in the row above the spots.
-
For more
detail select "View" from the Memory Menu
to display all individual measurements. The last
measurement values are displayed first. Scroll using
the up and down buttons.
Step
4 - Change the current SPOT or BATCH location.
To move onto
the next SPOT location select "New Spot" from
the gage menu. (Shortcut: press "+").
To move onto
the next BATCH location (begin a new PA2 analysis) select
"Memory -> New PA2 -> New" from the
gage menu..
Step
5 - Review results.
There are 4
ways to look at the results...
-
Upload
readings to PosiTector.net.
All PA2 information is presented including gage
serial number, PA2 parameters, individual measurements
with date and time stamp, calculated averages, and
Pass/Fail conclusions. View, annotate, and delete individual readings, and print reports. See below image for results
from a sample report.
-
View and
print reports offline using USB
Mass Storage. All PA2 information is presented
including gage serial number, PA2 parameters, individual
measurements with date and time stamp, calculated
averages, and Pass/Fail conclusions. See below image
for results from a sample report.
-
Print to
the optional
Bluetooth printer. All PA2 information is presented
including gage serial number, PA2 parameters, individual
measurements with date and time stamp, calculated
averages, and Pass/Fail conclusions.
-
Legacy
Method : All stored measurements can be downloaded
to your PC using PosiSoft
software. PosiSoft allows entry of notes and
annotations, prints and displays basic charts and
histograms, saves measurements to your PC, and exports
measurements to a spreadsheet or documents.
NOTES:
- Select the "Delete" menu option to erase
a SPOT, a BATCH, or the entire contents of gage memory.
- Select the "Close" menu option to end
a PA2 analysis but keep all existing BATCHES, SPOTS
and measurements.
- A gage RESET will restore PA2 parameters back to
factory settings (see the instruction manual for
RESET details).
The PosiTector 6000's PA2 feature is intended to support
the PA2 Application Standard, not replace it. It is
the responsibility of the user to read and understand
the document, and to verify the accuracy of the gage's
conclusions.

SSPC was founded in 1950 as the Steel Structures Painting
Council, a non-profit professional society concerned
with the use of coatings to protect industrial steel
structures. In 1997, the name of the association was
changed to The Society
for Protective Coatings to better reflect the changing
nature of coatings technology and the ever-expanding
types of construction materials.
SSPC is the only non-profit association that is focused
on the protection and preservation of concrete, steel
and other industrial and marine structures and surfaces
through the use of high-performance industrial coatings.
SSPC is the leading source of information on surface
preparation, coating selection, coating application,
environmental regulations, and health and safety issues
that affect the protective coatings industry.
SSPC standards, guides, specifications, and other
technical documents are developed in accordance with
voluntary consensus procedures established by SSPC
Technical Committees. They are intended to represent
a balance of interests, and are believed to represent
good current practice. All documents are monitored
and revised as practices improve.

The SSPC organization has developed a series of Paint
Application Standards, Guides and Specifications. One
of them is entitled, "SSPC-PA
2, Measurement of Dry Coating Thickness With Magnetic
Gages". It is among the most widely used SSPC
standards. (click here to
purchase the document).
PA2 describes the procedures to measure the thickness
of a dry film of a nonmagnetic coating applied on a
magnetic substrate using commercially available magnetic
gages. However the described methodology can be used
by most non-destructive coating thickness gages including
those using eddy-current or ultrasonic principles.
The standard defines a procedure to determine if the
film thickness over an extended area conforms to the
minimum and the maximum levels specified. This procedure
may be modified when measuring dry film thickness on
overcoated surfaces.
 |
Contents
Definitions
Description
of Gages
Calibration,
Verification, Adjustment, and Measurement Procedures
Required
number of Measurements for conformance to a
thickness specification
Notes
on gage principles and various factors affecting
thickness measurement
Substrate
Roughness, Tacky Films, Curvature, etc.
Variations
in Thickness
PA2
was originally developed to analyze large,
flat surfaces. But concerns about other shapes
and sizes prompted the creation of appendices.
- Appendix
1 - Numerical Example of Average Thickness
Measurement.
- Appendix
2 - Examples of the Adjustment of Electronic
Gages using Shims
- Appendix
3 - Methods for Measuring Dry Film Thickness
on Steel Beams (Girders)
- Appendix
4
-similar to Appendix 3 except that the pieces
are painted in a shop.
- Appendix
5
-Method for Measuring Dry Film Thickness on
Coated Steel Test Panels.
- Appendix
6
-Method for Measuring Dry Film Thickness of
Thin Coatings on Coated Steel Test Panels that
Had Been Abrasive Blast Cleaned.
|
 |
Definitions
| Gage
Reading: |
A single reading at one point. |
| Spot
Measurement: |
(sub-batch) The average of at
least three gage readings made within a 4 cm
(1.5 inch) diameter circle. |
| Dry
Film Thickness: |
(batch) The average of at least
five spot measurements for every 10 m2 (100
ft2). i.e. the "average
of the averages". |
| Calibration: |
The controlled and documented
process of measuring traceable calibration
standards and verifying that the results
are within the stated accuracy of
the gage. Calibrations are typically performed
by the gage manufacturer or by a qualified laboratory
in a controlled environment using a documented
process. Click here to
see a full set of measuring definitions. |
| TYPE
1 Gage: |
A magnetic pull-off gage such
as the PosiTest
FM. |
| TYPE
2 Gage: |
An electronic gage that uses electronic
circuitry, such as the PosiTector
6000. |
There are
several ways to evaluate film thickness with the PosiTector
6000 ...
-
Take
a single measurement.
-
Calculate
the average of several measurements -- the PosiTector
6000's Statistics mode.
-
Analyze
a larger area using a statistical approach --
the PosiTector 6000's PA2 mode.
A large surface
usually cannot be characterized accurately by a single
measurement. Variations in substrate preparation, painting
technique, measuring technique, etc. require an analysis
of several measurements taken over a large portion
of the surface.
The PosiTector 6000's statistics
mode displays individual measurements AND maintains
a running average value. It also displays minimum
and maximum values.
PA2
The
SSPC organization has taken this a step further
by developing a specification for structural
steel painters. Known as PA2 this specification
has become widely used both inside and outside
the structural steel painting industry.
PA2 helps determine if film thickness over
an extended area conforms to user-specified
minimum and maximum levels. |
 |

|
It recommends
that a coated surface be sectioned into one
or more large test areas. A minimum of 15 individual
measurements are taken in 5 groups of 3 measurements
each. Suggestions are made as to the acceptability
of individual measurements, spot measurements
(the average of each group), and the calculated
average of the spot measurements. In simple
terms, PA2 suggests the average of a series
of average values be used to determine
if a target thickness has been met. |
Repeated gage
readings, even at points close together, often
differ due to small surface irregularities
of the coating and the substrate. Therefore
PA2 suggests a minimum of three (3) individual
gage readings be taken for each spot measurement
of the coating. These readings are to be taken
within a 4 cm (1.5 inch) diameter circle. |
SSPC-PA2 Example

Advanced (3) versions of the PosiTector
6000 have a PA2 mode that helps the user collect
measurements in conformance with the PA2 standard.
Then, based on user-set parameters that include the
target thickness, the gage indicates a pass or fail
condition.
How does
the gage do this?
The PosiTector
6000 has on-board memory into which all measurements
are stored. Individual measurements are grouped into "spots" (sub-batches).
A series of spots are grouped into "batches".
The film thickness result obtained from each batch,
the "average of the averages", is taken to
be the representative film thickness over the defined
area.
The gage continually monitors thickness results and
reports a pass/fail condition on the current spot and
on the batch as a whole.
Related Documents:
ASTM
D7091 “Standard Practice for Nondestructive
Measurement of Dry Film Thickness of Nonmagnetic
Coatings Applied to a Ferrous Base and Nonmagnetic,
Nonconductive Coatings Applied to Non-Ferrous Metals”
ISO
19840 “Paints and varnishes --
Corrosion protection of steel structures by protective
paint systems - Measurement of, and acceptance criteria
for, the thickness of dry films on rough surfaces”
PosiTector 6000 Advanced
models have a statistics mode. When
selected, the gage continuously updates running
average and standard deviation calculations. Maximum
and minimum values are also tracked while measurements
are being taken.
The last measurement can
be deleted with the (-) button,
while the (+) button clears
all statistics. |
|
Is it possible
for all the spots to PASS but for the batch to
FAIL?
Yes. The pass/fail
criteria for a spot is less stringent than for a batch.
PA2 allows spot values to under run the target thickness
by as much as 20%. The batch value cannot.
Is it possible
to change PASS/FAIL criteria after all measurements
have been taken?
Yes. Select
the "PA2"
menu option, adjust the individual parameters, then
select "Exit". All stored PA2 batches will
instantly be re-evaluated and a new PASS/FAIL conclusion
made.
Can I delete
a bad measurement?
Yes. Use the
(-) button to delete the last measurement taken. Use
the "View" menu option to delete any measurement,
spot, or batch. Use the "Delete"
menu option to delete spots, batches or the entire
contents of memory.
Must I specify
both a minimum and a maximum thickness?
With many
of the coatings used today it is a good practice to
specify both. This range is usually obtained from the
coating manufacturer's technical data sheet. But economic
factors along with flow characteristics of the coating
usually control excessive film build, so it is common
to specify only a minimum thickness. When only a single
target value is available, set the MIN to that value
and set the MAX to 0.
|