PosiTector® 6000
PA2
Instruction
Manual Addendum
This addendum to instruction manual
v6.0/M is intended to provide support for the SSPC-PA2
feature found in all PosiTector 6000 memory(3)
models with serial numbers greater than 600000.
Step
1 - Begin a PA2 analysis.
| Select the Memory option from
the gage menu as shown. |
Next, select the PA2 option
from the Memory menu |
|
|
Review the 6 parameters.
To
adjust a parameter, use the (-) and (+) buttons
to move the highlighting bar to the appropriate
parameter, press the middle button, then
adjust that parameter as shown. |
 |
The
6 adjustable parameters are...
MIN:
the specified minimum thickness value
for the coating system. If only one target thickness
is specified, set the MIN to that value.
Adjust
the displayed value down (-) or
up (+). Alternatively, measure
a coating with a thickness close to the required
value and make final adjustments with the buttons.
Press
the middle button to save and exit. |
 |
MAX:
the specified maximum thickness value
for the structure. This value should be that
thickness above which adverse performance characteristics
appear, such as mudcracking, embrittlement, solvent
retention, loss of adhesion, cohesive splitting,
etc.
If a
maximum thickness value is not specified, set
the MAX value to 0.
Adjust
the displayed value down (-) or
up (+). Alternatively, measure
a coating with a thickness close to the required
value and make final adjustments with the buttons.
Press
the middle button to save and exit. |
 |
#/spot:
the minimum number of individual gage readings
required per spot. PA2 suggests a minimum
of 3 readings be taken.
Adjust
the displayed value down (-) or
up (+).
Press
the middle button to save and exit. |
 |
spots/batch:
the minimum number of spots (sub-batches) required
per batch. PA2 suggests a minimum of 5 spots
be measured.
Adjust the displayed
value down (-) or up (+).
Press the middle button
to save and exit.. |
|
In any measurement
there is a certain level of uncertainty. Two people
using the same gage will not necessarily record the
exact same number for a given spot measurement using
the same 4cm (1.5inch) diameter circle. To allow for
this natural fluctuation, an individual spot measurement
is permitted to be below the specified minimum thickness
as long as other spots in the batch area are high enough
to make the average thickness meet or exceed the specified
minimum thickness. Similar reasoning applies to maximum
thickness.
%
MIN:
PA2 specifies that no single spot measurement
be less than 80% of the specified MIN thickness.
The PosiTector 6000 allows this value to be changed
if desired.
Adjust the displayed
value down (-) or up (+).
Press the middle button to save and exit. |
 |
%
MAX:
PA2 specifies that no single spot measurement
be more than 120% of the specified MAX thickness.
The PosiTector 6000 allows this value to be changed
if desired.
If the MAX value
above has been set to 0, this % calculation is
made to the MIN value.
Adjust the displayed
value down (-) or up (+).
Press the middle button to save and exit. |
 |
To accept
these values but make no change to the operating
state of the gage, select Exit. This
allows for a handy review of current settings.
Changed parameters affect all existing and
future batches. The pass/fail standing of stored
batches can be changed this way.
To accept these values
and begin a new PA2 analysis, select NEW.
If a PA2 analysis was already begun before entering
this set-up, selecting NEW will close/save
the existing PA2 batch and begin a new one with
a batch name containing the next higher number. |
|
If NEW is
selected, a PA2 icon will appear on the gage's
display as shown.
Step
3 - Take measurements.
Begin taking
measurements in the first SPOT location. With each
measurement the following occurs...
-
The
green LED blinks, the gage BEEPS twice, and the
resultant measurement value is displayed in large
numerals at the bottom of the display.
-
The
SPOT information is updated
- "n" is total number
of individual measurements taken in the current
spot so far.
- "x" is the average
of all measurements in the spot. This is the
PA2 spot measurement value.
-a PASS or FAIL determination
is made for the SPOT based on the user-specified
parameters
-
The
BATCH information is updated
- "s" is the total
number of SPOTS in the BATCH
- "x" is the average
of the spot values i.e. the "average
of the averages".
- a PASS or FAIL determination
is made for the BATCH.
Step
4 - Change the current SPOT or BATCH location.
To move onto
the next SPOT location select "New Spot" from
the gage menu. (Shortcut: press "+").
To move onto
the next BATCH location (begin a new PA2 analysis)
select "Memory -> PA2 -> New" from
the gage menu..
Step
5 - Review results.
There are
4 ways to look at the results...
-
If all
PA2 criteria from Step
2 have been met, the word PASS will appear
beside the Batch number, i.e. "B1 PASS".
-
For
more detail select
"View" from the Memory Menu to display
all individual measurements. The last measurement
values are displayed first. Scroll using the
(-) and (+) buttons. Hold the (-) or (+) button
down for 1 second for fast scrolling.
--To replace a measurement, position the "+" symbol
beside the measurement. Then take another measurement
with the probe.
--To delete a measurement or to exit the View
screen, press the middle button.
-
All
stored measurements can be downloaded to your
PC using the included PosiSoft
software. PosiSoft allows entry of notes
and annotations, it prints and displays basic
charts and histograms, it saves measurements
to your PC, and it exports measurements to a
spreadsheet or documents.
-
Preferred
Method:
Print to the optional
IR printer or print to a PC's default printer
using the included PosiSoft
software. All PA2 information is presented
including gage serial number, PA2 parameters,
individual measurements with date and time stamp,
calculated averages, and PASS/FAIL conclusions.
This is the best way to ensure that the PA2 specification
has been satisfied. See sample .
NOTES:
- Select the "Delete" menu option to
erase a SPOT, a BATCH, or the entire contents of
gage memory. PA2 parameters are retained.
- Select the "Close" menu option to end
a PA2 analysis but keep all existing BATCHES, SPOTS
and measurements.
- Calibration adjustments cannot be made while PA2
measurements are stored in gage memory.
- A gage RESET will restore PA2 parameters back to
factory settings (see the instruction manual for
RESET details).
The PosiTector 6000's PA2 feature is intended to support
the PA2 Application Standard, not replace it. It is
the responsibility of the user to read and understand
the document, and to verify the accuracy of the gage's
conclusions.

SSPC was founded in 1950 as the Steel Structures Painting
Council, a non-profit professional society concerned
with the use of coatings to protect industrial steel
structures. In 1997, the name of the association was
changed to The Society
for Protective Coatings to better reflect the changing
nature of coatings technology and the ever-expanding
types of construction materials.
SSPC is the only non-profit association that is focused
on the protection and preservation of concrete, steel
and other industrial and marine structures and surfaces
through the use of high-performance industrial coatings.
SSPC is the leading source of information on surface
preparation, coating selection, coating application,
environmental regulations, and health and safety issues
that affect the protective coatings industry.
SSPC standards, guides, specifications, and other
technical documents are developed in accordance with
voluntary consensus procedures established by SSPC
Technical Committees. They are intended to represent
a balance of interests, and are believed to represent
good current practice. All documents are monitored
and revised as practices improve.

The SSPC organization has developed a series of Paint
Application Standards, Guides and Specifications. One
of them is entitled, "SSPC-PA
2, Measurement of Dry Coating Thickness With Magnetic
Gages". It is among the most widely used SSPC
standards. (click here to
purchase the document).
PA2 describes the procedures to measure the thickness
of a dry film of a nonmagnetic coating applied on a
magnetic substrate using commercially available magnetic
gages. However the described methodology can be used
by most non-destructive coating thickness gages including
those using eddy-current or ultrasonic principles.
The standard defines a procedure to determine if the
film thickness over an extended area conforms to the
minimum and the maximum levels specified. This procedure
may be modified when measuring dry film thickness on
overcoated surfaces.
 |
Contents
Definitions
Description
of Gages
Calibration,
Verification, Adjustment, and Measurement Procedures
Required
number of Measurements for conformance to a
thickness specification
Notes
on gage principles and various factors affecting
thickness measurement
Substrate
Roughness, Tacky Films, Curvature, etc.
Variations
in Thickness
PA2
was originally developed to analyze large,
flat surfaces. But concerns about other shapes
and sizes prompted the creation of appendices.
Appendix
1 - Numerical Example of Average Thickness
Measurement.
Appendix
2 - Examples of the Adjustment of Electronic
Gages using Shims
Appendix
3 - Methods for Measuring Dry Film Thickness
on Steel Beams (Girders)
Appendix
4
-similar to Appendix 3 except that the pieces
are painted in a shop.
Appendix
5
-Method for Measuring Dry Film Thickness on
Coated Steel Test Panels.
Appendix
6
-Method for Measuring Dry Film Thickness of
Thin Coatings on Coated Steel Test Panels that
Had Been Abrasive Blast Cleaned. |
 |
Definitions
| Gage
Reading: |
A single reading at one point. |
| Spot
Measurement: |
(sub-batch) The average of at
least three gage readings made within a 4 cm
(1.5 inch) diameter circle. |
| Dry
Film Thickness: |
(batch) The average of at least
five spot measurements for every 10 m2 (100
ft2). i.e. the "average
of the averages". |
| Calibration: |
The controlled and documented
process of measuring traceable calibration
standards and verifying that the results
are within the stated accuracy of
the gage. Calibrations are typically performed
by the gage manufacturer or by a qualified laboratory
in a controlled environment using a documented
process. Click here to
see a full set of measuring definitions. |
| TYPE
1 Gage: |
A magnetic pull-off gage such
as the PosiTest
FM. |
| TYPE
2 Gage: |
An electronic gage that uses electronic
circuitry, such as the PosiTector
6000. |
There are
several ways to evaluate film thickness with the PosiTector
6000 ...
-
Take
a single measurement.
-
Calculate
the average of several measurements -- the PosiTector
6000's Statistics mode.
-
Analyze
a larger area using a statistical approach --
the PosiTector 6000's PA2 mode.
A large surface
usually cannot be characterized accurately by a single
measurement. Variations in substrate preparation, painting
technique, measuring technique, etc. require an analysis
of several measurements taken over a large portion
of the surface.
The PosiTector 6000's statistics
mode displays individual measurements AND maintains
a running average value. It also displays minimum
and maximum values.
PA2
The
SSPC organization has taken this a step further
by developing a specification for structural
steel painters. Known as PA2 this specification
has become widely used both inside and outside
the structural steel painting industry.
PA2 helps determine if film thickness over
an extended area conforms to user-specified
minimum and maximum levels. |
 |

|
It recommends
that a coated surface be sectioned into one
or more large test areas. A minimum of 15 individual
measurements are taken in 5 groups of 3 measurements
each. Suggestions are made as to the acceptability
of individual measurements, spot measurements
(the average of each group), and the calculated
average of the spot measurements. In simple
terms, PA2 suggests the average of a series
of average values be used to determine
if a target thickness has been met. |
Repeated gage
readings, even at points close together, often
differ due to small surface irregularities
of the coating and the substrate. Therefore
PA2 suggests a minimum of three (3) individual
gage readings be taken for each spot measurement
of the coating. These readings are to be taken
within a 4 cm (1.5 inch) diameter circle. |
SSPC-PA2 Example

Memory (3) versions of the PosiTector
6000 have a PA2 mode that helps the user collect
measurements in conformance with the PA2 standard.
Then, based on user-set parameters that include the
target thickness, the gage indicates a pass or fail
condition.
How does
the gage do this?
The PosiTector
6000 has on-board memory into which all measurements
are stored. Individual measurements are grouped into "spots" (sub-batches).
A series of spots are grouped into "batches".
The film thickness result obtained from each batch,
the "average of the averages", is taken to
be the representative film thickness over the defined
area.
The gage continually monitors thickness results and
reports a pass/fail condition on the current spot and
on the batch as a whole.
Stored results can be viewed on the display, modified
or deleted. A formatted printout can be sent to the
optional IR
printer. Or all results can be downloaded to a
PC using the included PosiSoft
software and USB cable.
Related Documents:
ASTM
D7091-05 “Standard Practice for Nondestructive
Measurement of Dry Film Thickness of Nonmagnetic
Coatings Applied to a Ferrous Base and Nonmagnetic,
Nonconductive Coatings Applied to Non-Ferrous Metals”
ISO
19840:2004 “Paints and varnishes --
Corrosion protection of steel structures by protective
paint systems -- Measurement of, and acceptance criteria
for, the thickness of dry films on rough surfaces”
PosiTector 6000 Standard(2)
and Memory(3) gages have a statistics mode.
When selected, the gage continuously updates
running average and standard deviation calculations.
Maximum and minimum values are also tracked
while measurements are being taken.
The last measurement can
be deleted with the (-) button,
while the (+) button clears
all statistics. |
|
Is it possible
for all the spots to PASS but for the batch to
FAIL?
Yes. The pass/fail
criteria for a spot is less stringent than for a batch.
PA2 allows spot values to under run the target thickness
by as much as 20%. The batch value cannot.
Is it possible
to change PASS/FAIL criteria after all measurements
have been taken?
Yes. Select
the "PA2"
menu option, adjust the individual parameters, then
select "Exit". All stored PA2 batches will
instantly be re-evaluated and a new PASS/FAIL conclusion
made.
Can I delete
a bad measurement?
Yes. Use the
(-) button to delete the last measurement taken. Use
the "View" menu option to delete any measurement,
spot, or batch. Use the "Delete"
menu option to delete spots, batches or the entire
contents of memory.
NOTE: the "View"
menu option will also allow you to replace measurements.
Must I specify
both a minimum and a maximum thickness?
With many
of the coatings used today it is a good practice to
specify both. This range is usually obtained from the
coating manufacturer's technical data sheet. But economic
factors along with flow characteristics of the coating
usually control excessive film build, so it is common
to specify only a minimum thickness. When only a single
target value is available, set the MIN to that value
and set the MAX to 0.
|